Iso 80601-2-12 Test Conditions For Acoustic Tests Test Lung

Standard:ISO 80601-2-12-Table 201.102I:E Ratio:1:2
BAP:10 HPa (10 CmH2O)Set Rate:10/20/30 Breaths/min
Highlight:iso 80601-2-12 test conditions

acoustic test conditions

lung test conditions

  • Product Details
  • Inquiry

Product Description

ISO 80601-2-12 Test conditions for acoustic tests -test lung

 

Adjustable parameterTest condition
For a ventilator intended to provide tidal volume
VT ≥ 300 ml300 ml ≥ VT ≥ 50 mlVT ≤ 50 ml
Tidal volume, VTa500 ml150 ml30 ml
Set rate10 breaths/min20 breaths/min30 breaths/min
I:E ratio1:21:21:2
BAP10 hPa (10cmH2O)10 hPa (10cmH2O)10 hPa (10cmH2O)
Linear resistance,Rb5 hPa·(l/s) −1 ±10%20hPa·(l/s)−1 ±10%50hPa·(l/s)−1 ±10%
Isothermal Compliance, Cb50 ml·hPa −1 ±10 %20 ml·hPa −1 ±10 %1 ml·hPa −1 ±10 %
ModelKP-601250KP-601220KP-601201
a VT is measured by means of a pressure sensor at the test lung, where V T = C x (P max ), and VT is the volume delivered to the test lung

C is the isothermal compliance of the test lung

Pmax is the maximum pressure measured in the test lung

b The accuracy for C and R applies over the ranges of the measured parameters.

 