IEC 60601-2-52 Medical beds side rail strength and latch reliability tester
According to standard IEC 60601-2-52-cl.201.9.8.3.3.3Through PLC program control, 7 inch touch screen display, the use of electric cylinder to apply the test required force value, through high-precision sensor automatic data acquisition, can be arbitrarily set the test load value, load application time, test force peak acceleration capture function, maintain force stability function, reach the set time after the automatic cycle test function. After the test reaches the number set by the user, the device automatically stops and gives an alarm.
Main technical parameters:
1. Bed height: maximum 1200mm, width 1700mm;
2. Vertical force: 0-750N (down pressure);
3. Side force: 0-500N (push and pull);
4. Force sensor: 1000N, accuracy 0.3%;
5. Force sensor: number 3 groups;
6. Servo electric cylinder: 3 (400W servo motor);
7. Servo electric cylinder stroke: 0– 200mm;
8. Test fixture :3 sets;
9. Equipment size: 1200X2400X2000mm;
10. Equipment frame: aluminum profile or square pipe assembly.